FDAClass IINovember 28, 2012

KING LAD Standard ClearSeal SINGLE USE recall

KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 20 devices/case Size 2 / Infant ClearSeal Flexible King LAD¿ Patient Size: 10-20kg…

Why was KING LAD Standard ClearSeal SINGLE USE recalled?

The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF: LAD-902, Lot #K120414

Product
KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 20 devices/case Size 2 / Infant ClearSeal Flexible King LAD¿ Patient Size: 10-20kg…
Recalling firm
King Systems Corp.
Quantity
263 cases of product. (2,630 devices)
Distribution
US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
Recall date
November 28, 2012
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls