RAPIDPOINT 500 Blood Gas System recall
RAPIDPOINT 500 Blood Gas System
Why was RAPIDPOINT 500 Blood Gas System recalled?
There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Material No. 10492730, 10696855, 10696857, & 10697306; UDI-DI:00630414589169, 00630414590851, 00630414590868, & 00630414590844; All Units and Software Versions.
- Product
- RAPIDPOINT 500 Blood Gas System
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 12,363 units
- Distribution
- Domestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Republic, Costa Rica, Cote dIvoire, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Fiji, Finland, France, French, Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mexico, Moldova, Montenegro, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestinian Territory, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom Great Britain, Uruguay, & Vietnam.
- Recall date
- December 6, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)