Omnipod 5 Automated Insulin Delivery System Product recall
Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic…
Why was Omnipod 5 Automated Insulin Delivery System Product recalled?
The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
ALL serial numbers/ controller lots are currently in scope of this recall
- Product
- Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic…
- Recalling firm
- Insulet Corporation
- Quantity
- 129,086
- Distribution
- Domestic distribution nationwide.
- Recall date
- December 14, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)