FDAClass IIDecember 14, 2022

Flow-c Anesthesia System recall

Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

Why was Flow-c Anesthesia System recalled?

The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Flow-c UDI-DI: 7325710010457 Lot/Serial Numbers 4003 4777 4783 5024 5025 5026 5027 5028 5043 5044 5045 5046 5047 5048 5049 5050 5078 Flow-e UDI-DI: 7325710009765 Lot/Serial Numbers 50012 50013 50014 50015 50016 50017 50030 50031 50033 50034 50035 50036 50037 50038 50039 50040 50041 50042 50043 50046 50047 50048 50051 50052 50053 50054 50055 50056 50057 50058 50059 50060 50061 50062 50069 50070 50071 50072 50073 50074 50084 50085 50086 50087 50088 50089 50090 50091 50092 50093 50094 50096 50097 50108 50109 50110 50111 50132

Product
Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)
Recalling firm
Getinge Usa Sales Inc
Quantity
65 US
Distribution
US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.
Recall date
December 14, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)