LineSider Spinal System 5.5/6.0mm recall
LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation…
Why was LineSider Spinal System 5.5/6.0mm recalled?
Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
a. 5.5mm x 35mm, Part Number: LS-N6MG5535, UDI: (01)00818613028898(10)02210093, lot Number: 2210093; UDI: (01)00818613028898(10)10200053, Lot Number: 10200053. b. 5.5mm x 40mm, Part Number: LS-N6MG5540, UDI: (01)00818613028904(10)02210094, lot Number: 2210094; UDI: (01)00818613028904(10)10200054, Lot Number: 10200054. c. 5.5mm x 45mm, Part Number: LS-N6MG5545, UDI: (01)00818613028911(10)02210095, lot Number: 2210095; UDI: (01)00818613028911(10)10200055, Lot Number: 10200055. d. 5.5mm x 50mm, Part Number: LS-N6MG5550, UDI: (01)00818613028928(10)02210096, lot Number: 2210096; UDI: (01)00818613028928(10)10200056, Lot Number: 10200056. e. 6.5mm x 35mm, Part Number: LS-N6MG6535, UDI: (01)00818613028973(10)02210097, lot Number: 2210097; UDI: (01)00818613028973(10)10200057, Lot Number: 10200057; UDI: (01)00818613028973(10)10200057-1, Lot Number: 10200057-1. f. 6.5mm x 40mm, Part Number: LS-N6MG6540, UDI: (01)00818613028980(10)02210098, lot Number: 2210098; UDI: (01)00818613028980(10)10200058, Lot Number: 10200058; UDI: (01)00818613028980(10)10200058-1, Lot Number: 10200058-1. g. 6.5mm x 45mm, Part Number: LS-N6MG6545, UDI: (01)00818613028997(10)02210099, lot Number: 2210099; UDI: (01)00818613028997(10)10200059, Lot Number: 10200059; UDI: (01)00818613028997(10)10200059-1, Lot Number: 10200059-1. h. 6.5mm x 50mm, Part Number: LS-N6MG6550, UDI: (01)008181613029000(10)02210100, lot Number: 2210100; UDI: (01)00818613029000(10)10200060, Lot Number: 10200060; UDI: (01)00818613029000(10)10200060-1, Lot Number: 10200060-1. i. 6.5mm x 55mm, Part Number: LS-N6MG6555, UDI: (01)00818613029017(10)02210101, lot Number: 2210101; UDI: (01)00818613029017(10)10200061, Lot Number: 10200061. j. 7.5mm x 35mm, Part Number: LS-N6MG7535, UDI: (01)00818613026528(10)02210102, lot Number: 2210102; UDI: (01)00818613026528(10)10200062, Lot Number: 10200062. k. 7.5mm x 40mm, Part Number: LS-N6MG7540, UDI: (01)00818613026535(10)02210103, lot Number: 2210103; UDI: (01)00818613026535(10)10200063, Lot Number: 10200063; UDI: (01)00818613026535(10)06210309, Lot Number: 6210309. l. 7.5mm x 45mm, Part Number: LS-N6MG7545, UDI: (01)00818613026542(10)02210104, lot Number: 2210104; UDI: (01)00818613026542(10)10200064, Lot Number: 10200064; UDI: (01)00818613026542(10)10200064-1, Lot Number: 10200064-1. m. 7.5mm x 50mm, Part Number: LS-N6MG7550, UDI: (01)00818613026559(10)02210105, lot Number: 2210105; UDI: (01)00818613026559(10)10200065, Lot Number: 10200065; UDI: (01)00818613026559(10)10200065-1, Lot Number: 10200065-1. n. 7.5mm x 55mm, Part Number: LS-N6MG7555, UDI: (01)00818613026566(10)02210106, lot Number: 2210106; UDI: (01)00818613026566(10)10200066, Lot Number: 10200066. o. 8.5mm x 35mm, Part Number: LS-N6MG8535, UDI: (01)00818613026597(10)02210107, lot Number: 2210107; UDI: (01)00818613026597(10)10200067, Lot Number: 10200067. p. 8.5mm x 40mm, Part Number: LS-N6MG8540, UDI: (01)00818613026603(10)02210108, lot Number: 2210108; UDI: (01)00818613026603(10)10200068, Lot Number: 10200068. q. 8.5mm x 45mm, Part Number: LS-N6MG8545, UDI: (01)00818613026610(10)02210109, lot Number: 2210109; UDI: (01)00818613026610(10)10200069, Lot Number: 10200069. r. 8.5mm x 50mm, Part Number: LS-N6MG8550, UDI: (01)00818613026627(10)02210110, lot Number: 2210110; UDI: (01)00818613026627(10)10200070, Lot Number: 10200070. s. 8.5mm x 55mm, Part Number: LS-N6MG8555, UDI: (01)00818613026634(10)02210111, lot Number: 2210111; UDI: (01)00818613026634(10)10200071, Lot Number: 10200071. t. 9.5mm x 35mm, Part Number: LS-N6MG9535, UDI: (01)00818613026658(10)02210112, lot Number: 2210112; UDI: (01)00818613026658(10)10200072, Lot Number: 10200072. u. 9.5mm x 40mm, Part Number: LS-N6MG9540, UDI: (01)00818613026665(10)02210113, lot Number: 2210113; UDI: (01)00818613026665(10)10200073, Lot Number: 10200073. v. 9.5mm x 45mm, Part Number: LS-N6MG9545, UDI: (01)00818613026672(10)02210114, lot Number: 2210114; UDI: (01)00818613026672(10)10200074, Lot Number: 10200074. w. 9.5mm x 50mm, Part Number: LS-N6MG9550, UDI: (01)00818613026689(10)02210115, lot Number: 2210115; UDI: (01)00818613026689(10)10200075, Lot Number: 10200075. x. 9.5mm x 55mm, Part Number: LS-N6MG9555, UDI: (01)00818613026696(10)02210116, lot Number: 2210116; UDI: (01)00818613026696(10)10200076, Lot Number: 10200076.
- Product
- LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation…
- Recalling firm
- Integrity Implants Inc.
- Quantity
- 1,369 screws
- Distribution
- Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
- Recall date
- December 14, 2022
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)