da Vinci SP Surgical systems with the following recall: Software Issue
da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/…
Why was da Vinci SP Surgical systems with the following recalled?
Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may require surgeon to resolve an injury or need moderate intraoperative intervention. Also, use may potentially cause user frustration or a minor delay to troubleshoot.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All da Vinci SP Surgical systems with software version OS4 v4.0.0 (P4_B327)
- Product
- da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/…
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 46
- Distribution
- US Nationwide distribution in the states of Arizona, California, Connecticut, Florida, Illinois, Kansas, Kentucky, Massachusetts, Minnesota, Missouri, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Virginia, Washington, and Wisconsin.
- Recall date
- January 5, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)