FDAClass IIDecember 29, 2021

Philips Allura Xper FD series with Software Version recall

Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging applications…

Why was Philips Allura Xper FD series with Software Version recalled?

Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

US Serial Number/UDI: 1 (01)00884838054202(21)1; 527 (01)00884838059054(21)527; 408 (01)00884838054202(21)408; 306 (01)00884838054202(21)306; 61 (01)00884838059030(21)61 OUS: Model Product S/N UDI: 722006 Allura Xper FD20 1997 (01)00884838059191(10)1997; 722010 Allura Xper FD10 1177 (01)00884838059030(21)1177; 722010 Allura Xper FD10 1178 (01)00884838059030(21)1178; 722013 Allura Xper FD20 Biplane 544;; 722010 Allura Xper FD10 514 (01)00884838059030(21)514; 722013 Allura Xper FD20 Biplane 271; 722025 Allura Xper FD20 Biplane OR Table 6; 722005 Allura Xper FD10/10 519; 722008 Allura Xper FD20 Biplane 531; 722012 Allura Xper FD20 1901 (01)00884838059054(21)1901; 722012 Allura Xper FD20 2505 (01)00884838059054(21)2505; 722013 Allura Xper FD20 Biplane 137; 722026 Allura Xper FD10 439 (01)00884838054189(21)439; 722134 Field ext. Xper vascular systems R7.6 98736; 722027 Allura Xper FD10/10 172 (01)00884838054196(21)172;

Product
Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging applications…
Recalling firm
Philips North America Llc
Quantity
USA: 5 systems R.O.W.: 15 systems
Distribution
Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.
Recall date
December 29, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls