Smith & Nephew Taylor Spatial Frame Website software recall: Software Issue
Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral…
Why was Smith & Nephew Taylor Spatial Frame Website software recalled?
In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Version 5.2.5 and 5.2.6
- Product
- Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral…
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- N/A
- Distribution
- Worldwide
- Recall date
- January 31, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)