FDAClass IIDecember 16, 2015

One Step Button recall

One Step Button, Low Profile Initial Placement PEG Kit, gastro-enterostomy tube.

Why was One Step Button recalled?

Boston Scientific Corp. initiated a voluntary recall of specific lots of the EndoVive One Step Button, including the Decompression Tube and Low Profile Replacement Button devices because they may contain decompression tube adapters with a protrusion that could impede its ability to be introduced or removed from the PEG Button.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Material Number (UPN) M00563020, Lot # 18184947, Exp. April 30, 2017; Material Number M00563080, Lot # 18179867, Exp. April 30, 2017; & Lot # 18212410 , Exp. May 31, 2017; Material Number M00568510, Lot # 18296155, Exp. May 31, 2018.

Product
One Step Button, Low Profile Initial Placement PEG Kit, gastro-enterostomy tube.
Recalling firm
Boston Scientific Corp
Quantity
22 units
Distribution
US Distribution to the states of : IN, NV, CA, PA, CT, TN, NY, WI, and WA, and Internationally to Ireland.
Recall date
December 16, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls