Specialty Audible Torque Wrench recall
SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible…
Why was Specialty Audible Torque Wrench recalled?
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Class I device, 510(K) exempt Catalog Number: IS2217XLP Lot Number: 2576, 2756
- Product
- SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible…
- Recalling firm
- Stryker Spine
- Quantity
- 28 units
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
- Recall date
- November 28, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)