CODMAN Disposable Perforator 9mm. Cranial Perforator recall
CODMAN Disposable Perforator 9mm. Cranial Perforator.
Why was CODMAN Disposable Perforator 9mm. Cranial Perforator recalled?
Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model No. 261223; UDI: 10381780513605; Lot No. 5828553, 6264478, 6264481, 6264482, 7230672, 7230673, 7230674, 7230675, 7261971, 7268008, 7268009, 7268010, 7268011, 7268012, 7268013, 7268014, 7268015, 7268016, 7268017, 7268018, 7268019, 7282954, 7282955, 7293422, 7293423, 7293424, 7293425, 7293426, 7293427, 7293428, 7300667, 7300668, 7300669, 7300670, 7300672, 7300673, 7300674, 7371082, 7371083, 7379652, 7379922, 7379923, 7379924, 7379926, 7379928, 7379660, 7379925, 7379927, 7450559, 7450560, 7450561, 7450562, 7450563, 7450564, 7450565, 7450566, 7451273, 7451274, 7451275, 7451276, 7451277, 7451278, 7451279; Expiration Date: 2024-03-31 to 2029-07-31.
- Product
- CODMAN Disposable Perforator 9mm. Cranial Perforator.
- Recalling firm
- Integra LifeSciences Corp. (NeuroSciences)
- Quantity
- 5,109 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czechia, Ecuador, France, French Guiana, Germany, Greece, India, Ireland, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Morrocco, Netherlands, New Zealand, Palestine, State of, Philippines, Portugal, Puerto Rico, Qatar, Republic of North Macedonia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan , Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Vietnam.
- Recall date
- November 12, 2025
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)