2.5 mm LifePort Endotracheal Tube Adapters 10-pack recall
2.5 mm LifePort Endotracheal Tube Adapters 10-pack, Catalog: 9025; and 2.5 mm LifePort Endotracheal Tube Adaptors inside Patient Circuit Kits…
Why was 2.5 mm LifePort Endotracheal Tube Adapters 10-pack recalled?
Some adaptors, for use with high frequency ventilators, were packaged improperly, and may deform at the tip, which may impair the ability to pass a suction catheter or may affect normal ventilator performance, which may lead to hypercarbia, hypoxia, and gas trapping.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog/UDI/Lot: 9025/10616120000034/22042320 ; and 2.5 mm LifePort Endotracheal Tube Adaptors, Lot: 22042320, inside 938/10616120000089, 00616120000082/22052381, 22052407, 22062431, 22062458, 22062478
- Product
- 2.5 mm LifePort Endotracheal Tube Adapters 10-pack, Catalog: 9025; and 2.5 mm LifePort Endotracheal Tube Adaptors inside Patient Circuit Kits…
- Recalling firm
- Bunnell, Inc.
- Quantity
- 1606 Assemblies
- Distribution
- US Nationwide Distribution: PA, NY, AR, AZ, TN, SC, MA, FL, CA, TX, MD, VA, NV, IL, IN, IA, HI, KY, LA, WI, GA, OH, MI, MO, MN, NJ, MS, NC, ND, NM, UT, OK, WA, RI; OUS: Canada
- Recall date
- December 14, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)