Stryker Neptune SafeAir Smoke Evacuation Pencil Non recall
Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including…
Why was Stryker Neptune SafeAir Smoke Evacuation Pencil Non recalled?
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GTIN: 07613327117196 Lot Numbers: 53867671 54409847 59053476 64855502 66494932
- Product
- Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including…
- Recalling firm
- Stryker Corporation
- Quantity
- 21,490 eaches
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
- Recall date
- November 20, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)