Curaplex manual resuscitators with integrated manometer recall

Curaplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep Valve & Mask, REF # 2442-BVMPAD; 2. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & Mask, REF # 2442-BVMPCH; 3. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter, Infant and Neonate Mask, REF # 2442-BVMCIN; 4. Curaplex Infant Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & 2 Masks, REF # 2442-BVMPIN; 5. Curaplex Small Adult Pediatric Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter, PEEP Valve and 2 Masks, REF # 2442-BVMPSA.

Why was Curaplex manual resuscitators with integrated manometer recalled?

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 00810071633126, Item # 2442-BVMCIN; UDI-DI: 00812277037050, Item # 2442-BVMPAD; UDI-DI: 00812277037074, Item # 2442-BVMPCH; UDI-DI: 00812277037081, Item # 2442-BVMPIN; UDI-DI: 00812277037067, Item # 2442-BVMPSA. All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.

Product
Curaplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep Valve & Mask, REF # 2442-BVMPAD; 2. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & Mask, REF # 2442-BVMPCH; 3. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter, Infant and Neonate Mask, REF # 2442-BVMCIN; 4. Curaplex Infant Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & 2 Masks, REF # 2442-BVMPIN; 5. Curaplex Small Adult Pediatric Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter, PEEP Valve and 2 Masks, REF # 2442-BVMPSA.
Recalling firm
SunMed Holdings, LLC
Quantity
N/A
Distribution
US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
Recall date
November 29, 2023
Classification
Class II
Status
Ongoing