Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation recall
Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting…
Why was Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation recalled?
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GTIN: 07613327117219 Lot Number: 64117291
- Product
- Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting…
- Recalling firm
- Stryker Corporation
- Quantity
- 2,620 eaches
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
- Recall date
- November 20, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)