FDAClass IIDecember 22, 2021

Smith&nephew Evos Medial Distal Femur Plate recall

smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201

Why was Smith&nephew Evos Medial Distal Femur Plate recalled?

Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Batch Number: 20HM10255, 20HM10256, 20HM10257, UDI

Product
smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201
Recalling firm
Smith & Nephew, Inc.
Quantity
35 units
Distribution
US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.
Recall date
December 22, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)