Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch recall
Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended…
Why was Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch recalled?
Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: Each 10884521009585 Case 20884521009582 Lot Numbers: 2216806064, 2214452064, 2212401164 2210219764, 2209525364, 2208816464 2206737964, 2206727564
- Product
- Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended…
- Recalling firm
- Cardinal Health
- Quantity
- N/A
- Distribution
- Nationwide
- Recall date
- December 14, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)