Infinix DP- i(Infinix-8000F) CAT-870B Catheterization recall
Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System…
Why was Infinix DP- i(Infinix-8000F) CAT-870B Catheterization recalled?
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model CAT-870B D053991 Serial No. Manufacture Date 99D08X2101 October 2008 99D0582039 August 2005 99D0762072 June 2007 99D0852090 May 2008 99D07Y2078 November 2007 99D0812081 January 2008 99D0582042 August 2005 99D0552033 May 2005 D551206 May 2005
- Product
- Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System…
- Recalling firm
- Canon Medical System, USA, INC.
- Quantity
- 9 systems
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.
- Recall date
- November 20, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)