Altrus Thermal Tissue Fusion Energy Sources recall
Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.
Why was Altrus Thermal Tissue Fusion Energy Sources recalled?
Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: 1) 15090103 & 15150005, 15150006, 15150007, 15150008, 15150009, 15150010, 15150011, 15150012, 15130004, 15250013, 15250014, 15250015, 15250016, 15250017, 15300018 & 15300020, 2) 15050001, 15300005 & 15300006
- Product
- Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.
- Recalling firm
- ConMed Corporation
- Quantity
- 9 devices
- Distribution
- Distributed in the states of CA, and TX and in the country of Mexico.
- Recall date
- December 9, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)