FDAClass IIDecember 14, 2022

IMMULITE and IMMULITE 1000 Total Testosterone Assay- In recall

IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers for the quantitative…

Why was IMMULITE and IMMULITE 1000 Total Testosterone Assay- In recalled?

Confirmed an average positive bias of 40% for testosterone concentrations of approximately 660 ng/dL (22.9 nmol/L) and above with the IMMULITE/IMMULITE 1000 Total Testosterone Assay when compared to the IMMULITE 2000/IMMULITE 2000 XPi Total Testosterone Assay. Two (2) lots of Total Testosterone assays for the IMMULITE/IMMULITE 1000, kit lots #515 and #516 were actively distributed to customers when the issue was discovered by Siemens, and they are subject to the correction or removal in this recall (867 total quantity of reagents distributed for each lot). The violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. In both males and female, testosterone results are not used in isolation but are interpreted in conjunction with clinical history and symptomology as well as with other laboratory testing (e.g. gonadotropins and other sex steroids). A delay in the diagnosis may be experienced and medical intervention, such as repeat testing may be required.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: (01)00630414964782(10)0515(17)20221130; (01)00630414964782(10)0516(17)20221231 Lot Numbers: 515, Exp Date: 2022-11-30 516, Exp Date: 2022-12-31

Product
IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers for the quantitative…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
867 units
Distribution
Worldwide Distribution: US (nationwide) to states of: AL, CA, FL, GA, IN, KY, NC, NE, OR, SD, TX, UT; and OUS (Foreign) to countries of: Canada, Mexico, Austria, Argentina, Austria, Bangladesh, Colombia, Brazil, Bulgaria, Costa¿Rica, Cyprus, Estonia, Germany, Greece, India, U.A.E., Italy, Kazakhstan, Latvia, P.R.¿China, Romania, Russian¿Fed., South Africa, Spain, United Kingdom, Uruguay and Singapore.¿¿¿¿
Recall date
December 14, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)