Two-stage and three-stage I.V. poles are shipped as an recall
Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable…
Why was Two-stage and three-stage I.V. poles are shipped as an recalled?
It was identified by a customer complaint that some I.V. poles were retracting/lowering from an extended position to the low height unexpectedly. An investigation was initiated which identified that the I.V. Latch Housing was nonconforming. potential hazards: " Suspended I.V. pole retracts unexpectedly " Suspended I.V. pole retracts unexpectedly with pump/bags attached
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Stryker 2-Stage and 3-Stage I.V. Poles distributed on beds, stretchers, and in service kits. This product field action is limited to I.V. Poles distributed between 01/07/2014 - 06/24/2014. The following I.V. Pole option numbers come affixed to beds and stretchers: Manufacturer Model Number and Product Description 2035-112-010; Head End Left, 2 Stage I.V. Pole 2035-113-011; Head End Right, 2 Stage I.V. Pole 0785-035-101; 26", 2 Stage IV Pole Assembly 0785-035-200; 26", 3 Stage IV Pole Assembly 0785-035-300; 30", 3 Stage IV Pole Assembly 0785-035-401; 30", 2 Stage IV Pole Assembly The following I.V. Pole option numbers ship as service kits. Manufacturer Service Kit Number and Product Description 1105-035-368; 30" IV POLE 2 STAGE HE PR 1105-035-367; 30" IV POLE 2 STAGE HE PL 1105-035-343; 30" IV POLE 2 STAGE FE PR 1105-035-341; 30" IV POLE, 2 STAGE HE, PR 1105-035-340; 30" IV POLE, 2 STAGE FE, PL 1105-035-338; 30" IV POLE, 2 STAGE FE, PL 1040-035-338; 2-STAGE IV POLEHEAD LT30" ZM 0785-035-340; 2-STAGE IV POLE,HEAD LT, 30" 0785-035-341; 2-STAGE IV POLEHEAD RT 30" 0785-035-343; 2-STAGE IV POLE,FOOT RT, 30" 1105-035-344; 30" IV POLE, 3 STAGE FE, PL 1105-035-342; 30" IV POLE, 3 STAGE HE, PL 1105-035-339; 30" IV POLE, 3 STAGE FE, PR 1105-035-369; 30" IV POLE, 3 STAGE HE, PL 1105-035-366; 30" IV POLE, 3 STAGE HE, PR 0785-035-339; 3-STAGE IV POLE,FOOT RT, 30" 0785-035-342; 3-STAGE IV POLE,HEAD LT, 30" 1105-035-644; 26" IV POLE, 3 STAGE FL 1105-035-642; 26" IV POLE, 3 STAGE HL 1105-035-639; 26" IV POLE 3 STAGE FR 1105-035-637; 26" IV POLE, 3 STAGE HR 0785-035-637; 3-STAGE IV POLE,HEAD RT, 26" 0785-035-639; 3-STAGE IV POLE,FOOT RT, 26" 0785-035-642; 3-STAGE IV POLE,HEAD LT, 26" 1105-035-643; 26" IV POLE, 2 STAGE FR 1105-035-641; 26" IV POLE, 2 STAGE HR 1105-035-640; 26" IV POLE, 2 STAGE FL 1105-035-638; 26" IV POLE, 2 STAGE HL 0785-035-638; 2-STAGE IV POLE,HEAD LT, 26" 0785-035-640; 2-STAGE IV POLE,FOOT LT, 26" 0785-035-641; 2-STAGE IV POLE,HEAD RT, 26" 0785-035-643; 2-STAGE IV POLE,FOOT RT, 26" 2035-112-000; IV POLE - 2 STAGE H/E 2035-113-000; DUAL H/E 2 STAGE IV POLE OPT
- Product
- Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable…
- Recalling firm
- Stryker Medical Division of Stryker Corporation
- Quantity
- 19,876
- Distribution
- Worldwide Distribution-US (nationwide) including DC and Guam, and the countries of New Zealand Canada Chile Hong Kong Colombia Argentina Brazil France Germany Italy Netherlands Spain Switzerland United Kingdom India Japan South Korea Mexico South Africa Poland Singapore.
- Recall date
- December 9, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)