Puritan Bennett 980 Series Ventilator recall
Puritan Bennett 980 Series Ventilator, 980S3ENDICU
Why was Puritan Bennett 980 Series Ventilator recalled?
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GTIN/UDI - 10884521171244 Serial Numbers: 35B1700406 35B1700449 35B1700467 35B1700469 35B1700471 35B1700474 35B1700481 35B1700526 35B1700558
- Product
- Puritan Bennett 980 Series Ventilator, 980S3ENDICU
- Recalling firm
- Covidien Llc
- Quantity
- 278 total
- Distribution
- Worldwide distribution - US Nationwide.
- Recall date
- December 15, 2021
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)