FDAClass IINovember 29, 2023

Tracheal fiberscope-airway management recall: Contamination

Tracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management.…

Why was Tracheal fiberscope-airway management recalled?

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number (UDI-DI): LF-DP (04953170136856), LF-GP (04953170061998), LF-TP (04953170317576).

Product
Tracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management.…
Recalling firm
Aizu Olympus Co., Ltd.
Quantity
65 units
Distribution
Worldwide - US Nationwide distribution.
Recall date
November 29, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)