Cardiosave Rescue IntraAortic Balloon Pump recall
Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83, 0998-00-0800-85…
Why was Cardiosave Rescue IntraAortic Balloon Pump recalled?
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-75, 10607567112312; 0998-UC-0800-75, 10607567112312; 0998-00-0800-83, 10607567108407; 0998-UC-0800-83, 10607567108407; 0998-00-0800-85, 10607567113449; 0998-UC-0800-85, 10607567113449.
- Product
- Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83, 0998-00-0800-85…
- Recalling firm
- Datascope Corp.
- Quantity
- 44 (US); 5 (OUS)
- Distribution
- Worldwide distribution - US Nationwide.
- Recall date
- December 22, 2021
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)