Covidien Sonicision Battery Charger-intended to charge recall
Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging…
Why was Covidien Sonicision Battery Charger-intended to charge recalled?
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: MCSB006939, MCSB006947 GTIN: 10884521592896
- Product
- Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging…
- Recalling firm
- Covidien Llc
- Quantity
- 2 units OUS
- Distribution
- Brazil
- Recall date
- December 15, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)