Bronchoscope - Ultrasonic: endoscopic real-time recall: Contamination
Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the…
Why was Bronchoscope - Ultrasonic: endoscopic real-time recalled?
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number (UDI-DI): BF-UC180F (04953170356360), BF-UC190F (04953170399831).
- Product
- Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the…
- Recalling firm
- Aizu Olympus Co., Ltd.
- Quantity
- 823 units
- Distribution
- Worldwide - US Nationwide distribution.
- Recall date
- November 29, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)