IRON assay recall
IRON assay, Reference Numbers 6K95-41 and 6K95-30
Why was IRON assay recalled?
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Reference Number 6K95-41: UDI-DI: 00380740010928; Lot Number: 30610Y600; Reference Number 6K95-30: UDI-DI: 00380740010911; Lot Numbers: 21058Y600, 21101Y600, 30918Y600
- Product
- IRON assay, Reference Numbers 6K95-41 and 6K95-30
- Recalling firm
- Sentinel CH SpA
- Quantity
- 4394 kits
- Distribution
- Worldwide - US Nationwide distribution.
- Recall date
- November 29, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)