Solea Models 2.0 and 3.0 Laser Surgical Instrument recall
Solea Models 2.0 and 3.0 Laser Surgical Instrument, Foot Pedal assembly SUB-00434 with foot pedal CUS-00226 Rev. K reworked
Why was Solea Models 2.0 and 3.0 Laser Surgical Instrument recalled?
It has been found that potential unintended laser activation can occur without foot pedal depression.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Solea Model 3.0: UDI-DI: 00850634007018; Serial Numbers: 0310516, 0310517, 0310518, 0310519, 0310520, 0310521, 0310522, 0310523, 0310524, 0310525, 0310526, 0310527, 0310528, 0310529, 0310530, 0310531, 0310532, 0310533, 0310534, 0310535, 0310536, 0310537, 0310538, 0310539, 0310540, 0310541, 0310542, 0310543, 0310317, 0310321, 0310322, 0300730; Solea Model 2.0: UDI-DI: 00850634007001; Serial Numbers: 0200822, 0200807, 0200810, 0200625, CDS2-U33W-1B-DE-15, 0200819, 0200401
- Product
- Solea Models 2.0 and 3.0 Laser Surgical Instrument, Foot Pedal assembly SUB-00434 with foot pedal CUS-00226 Rev. K reworked
- Recalling firm
- Convergent Dental
- Quantity
- 39 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, AL, CA, CT, FL, KY, ME, MI, MO, MT, NC, OH, OR, PA, TN, TX, UT, VA, WA, WI and the country of Canada.
- Recall date
- November 29, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)