FDAClass IINovember 2, 2016

Immulite 2000/immulite 2000xpi Systems CMM CMV Igm recall

IMMULITE 2000/IMMULITE 2000XPi Systems CMM CMV IgM Assay, REF/Catalog Number L2KCM2(D) - 200 tests (US), SMN 10370302 (US), UDI 00630414985725 (US)…

Why was Immulite 2000/immulite 2000xpi Systems CMM CMV Igm recalled?

Multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI 00630414985725/Catalog # L2KCM2(D): Kit lots D255, D256, D257, D258, D259, D260, D261, D262, D263, D264, D266; UDI 00630414961163/Catalog # L2KCM2: Kit lots 255, 256, 257, 258, 259, 260, 261, 262, 263, 264, 266, 267

Product
IMMULITE 2000/IMMULITE 2000XPi Systems CMM CMV IgM Assay, REF/Catalog Number L2KCM2(D) - 200 tests (US), SMN 10370302 (US), UDI 00630414985725 (US)…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
US: 326 kits; Foreign: 9633 kits
Distribution
Worldwide distribution- US (Nationwide) and countries of: United Arab Emirates, Afghanistan, Angola, Argentina, Australia, Bangladesh, Belgium, Bulgaria, Bahrain, Bolivia, Brazil, Belarus, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Dominican Republic, Algeria, Ecuador, Egypt, Spain, France, Great Britain, Greece, Guatemala, Honduras, Hungary, India, Iraq, Italy, Jordan, Canary Islands, Kyrgyzstan, Rep. Korea (S), Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macedonia, Mexico, Malaysia, Nicaragua, Netherlands, Norway, Nepal, Peru, Pakistan, Poland, Portugal, Paraguay, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, El Salvador, Syria, Turkmenistan, Tunisia, Turkey, Uruguay, Uzbekistan, Vietnam, Serbia,and South Africa.
Recall date
November 2, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)