Mobile Detector Holder - Stationary X-ray System recall
Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray…
Why was Mobile Detector Holder - Stationary X-ray System recalled?
If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
9890-010-88152 - Codes included in attached list. Devices were manufactured and distributed from March 2015 - March 2017.
- Product
- Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray…
- Recalling firm
- Philips Electronics North America Corporation
- Quantity
- 412
- Distribution
- 412 units affected.
- Recall date
- January 10, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)