FDAClass IINovember 22, 2023

AU/DxC AU Chemistry Transferrin recall

AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solution scatter light in proportion to their size, shape and concentration.…

Why was AU/DxC AU Chemistry Transferrin recalled?

The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 15099590010492, LOT: 2573, Expiration: 01 August 2024

Product
AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solution scatter light in proportion to their size, shape and concentration.…
Recalling firm
Beckman Coulter Biomedical Division
Quantity
2,754
Distribution
Worldwide - US Nationwide distribution including in the states of NY, CA, MS, GA, PA, SC, WA, TN, AL, AK, OR, MO, KS, FL, LA, OH, TX, NJ, MI, WV, MD, IL, RI, VA, NC, IN, OK, NV, NE, WY, KY, WI, ND, MA, NH, UT, CO, GU, PR, AR, AZ, ME and the countries of Canada, Netherlands, United Kingdom, Sweden, Germany, France, Switzerland, Tunisia, Australia, Belgium, Costa Rica, Uruguay, Turkey, South Africa, New Zealand, Italy, Thailand, Ghana, Russia, Hungary, Czech Republic, Slovakia, India, Austria, Poland, Romania, Ireland, Spain, Portugal, Mexico, Vietnam, Korea, Mozambique, Kazakhstan, China, Malaysia, Singapore, Kenya, Brazil, Namibia, Egypt, Panama.
Recall date
November 22, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)