Brand Name: Atlan Product Name: Atlan A350XL recall
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
Why was Brand Name: Atlan Product Name: Atlan A350XL recalled?
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model/Catalog Number: 8621600; UDI-DI Number: 04048675556183; Serial Numbers Distributed in the US: ASTA-0307; 0309; 0311; 0349; 0350; 0351; 0353; 0363; 0365; 0394; 0408; 0409 ASTB-0048; 0286; 0287; 0288; 0289; 0290; 0291; 0304; 0305; 0306; 0307; 0308; 0309; 0317; 0318; 0319; 0320; 0321; 0322 ASTC-0006; 0007; 0068; 0069; 0070; 0082; 0083; 0099; 0100; 0101; 0105; 0106; 0107; 0108; 0109; 0118; 0119; 0136; 0137; 0229; 0230; 0243; 0244; 0245 ASTD-0008; 0009; 0010; 0011; 0012; 0013; 0014; 0129; 0130; 0131; 0132; 0133; 0134; 0135; 0154; 0159; 0160; 0161; 0162; 0163; 0164; 0165; 0166; 0167; 0168; 0169; 0181; 0182; 0183; 0193; 0194; 0195; 0225; 0226; 0227; 0228; 0229; 0230; 0231; 0238; 0239; 0240; 0241; 0242; 0243; 0244; 0245 ASTE-0016; 0017; 0077; 0078; 0079; 0086; 0087; 0122; 0123; 0124; 0125; 0126; 0129; 0130; 0131; 0163; 0164; 0165; 0166; 0171; 0172; 0173; 0174; 0195; 0196; 0197; 0198; 0203; 0225; 0226; 0246; 0247; 0248; 0249; 0250; 0278; 0279; 0280; 0281; 0301; 0302; 0303; 0304; 0305; 0306; 0348; 0349; 0350; 0351; 0352; 0357; 0358; 0359; 0360; 0361; 0362 ASTF-0001; 0002; 0003; 0004; 0014; 0015; 0016; 0017; 0018; 0030; 0031; 0032; 0033; 0034; 0035; 0036; 0037; 0055; 0061; 0062; 0063; 0070; 0071; 0076; 0077; 0078; 0079; 0080; 0081; 0121; 0122; 0123; 0124; 0125; 0135; 0136; 0137; 0138; 0139; 0140; 0141; 0168; 0169; 0170; 0194; 0195; 0196; 0409; 0410; 0411; 0436; 0437; 0438; 0474; 0475; 0476; 0477; 0478 ASTH-0007; 0008; 0009; 0010; 0011; 0012; 0092; 0093; 0094; 0095; 0096; 0120; 0121; 0122; 0123; 0124; 0125; 0126; 0142; 0143; 0144; 0145; 0146; 0147; 0148; 0169; 0170; 0171; 0172; 0173; 0174; 0175; 0187; 0188; 0189; 0190; 0191; 0192; 0269; 0270; 0271; 0272; 0273; 0293; 0294; 0295; 0296; 0306; 0307; 0308; 0309; 0310; 0311; 0324; 0325; 0326; 0327; 0330; 0331; 0332; 0333; 0334; 0335; 0336; 0337; 0338; 0339; 0358; 0359; 0360; 0361; 0362; 0363; 0364; 0365; 0383; 0384; 0385; 0386; 0387; 0388; 0389; 0390; 0450; 0476; 0477; 0478; 0479; 0480; 0481; 0488; 0489; 0490; 0491 Serial Numbers Distributed OUS: ASTA-0253; 0289; 0290; 0327; 0362; 0404; 0405; 0426; 0427; 0428; 0429; 0430 ASTB-0003; 0005; 0007; 0028; 0029; 0030; 0031; 0032; 0033; 0075; 0076; 0077; 0079; 0080; 0081; 0082; 0083; 0084; 0134; 0195; 0206; 0219; 0220; 0237 ASTC-0001; 0002; 0003; 0004; 0026; 0138; 0163; 0180; 0181; 0194; ASTD-0015; 0016; 0050; 0061; 0062; 0063; 0064; 0065; 0066; 0203; 0204; 0205; 0265; 0266; 0267 ASTE-0034; 0035; 0057; 0058; 0059; 0060; 0062; 0085; 0108; 0127; 0128; 0175; 0202; 0223; 0224; 0300; 0325; 0326; 0327; 0328; 0329; 0353; 0354; 0355; 0356 ASTF-0051; 0052; 0053; 0054; 0112; 0113; 0435; 0471 ASTH-0039; 0040; 0045; 0046; 0081; 0082; 0176; 0193; 0194; 0204; 0217; 0218; 0219; 0225; 0226; 0243; 0244; 0245; 0246; 0255; 0256; 0382; 0403; 0404; 0405; 0406; 0434; 0435; 0436; 0437; 0451; 0452; 0453; 0 ASTJ-0063; 0064; 0065; 0066; 0067; 0068; 0089; 0090; 0091; 0092; 0145; 0146; 0147; 0171; 0237 ASTK-0002
- Product
- Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 455 units (310 units US, 145 OUS)
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
- Recall date
- November 20, 2024
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)