ReTurn 7500/7500 Aid recall
ReTurn 7500/7500 Aid, transfer
Why was ReTurn 7500/7500 Aid recalled?
During the period of April 2009 May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model # 7400, 7500 and 7322 (spare part) Serial # 373317697270501334, 373317697270501430, 073317697400500012, 073317697400500057
- Product
- ReTurn 7500/7500 Aid, transfer
- Recalling firm
- Handicare USA, Inc.
- Quantity
- 29 units , plus 47 wing handles were shipped as spare parts
- Distribution
- US Distribution to the states of : NY, NJ, NV, NE, IL, CT, PA and ME
- Recall date
- November 2, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)