FDAClass IINovember 25, 2015

Fresenius 132 Gallon GranuFlo Dissolution Unit - P/N recall: Failed Dissolution

Fresenius 132 Gallon GranuFlo Dissolution Unit - P/N G047-80101. For Use in the Preparation of acid concentrate for hemodialysis.

Why was Fresenius 132 Gallon GranuFlo Dissolution Unit - P/N recalled?

Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors.

Product
Fresenius 132 Gallon GranuFlo Dissolution Unit - P/N G047-80101. For Use in the Preparation of acid concentrate for hemodialysis.
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Quantity
918 units
Distribution
Nationwide Distribution.
Recall date
November 25, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls