FDAClass IINovember 22, 2023

Brilliance CT Big Bore with software version recall

Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and therapy.

Why was Brilliance CT Big Bore with software version recalled?

For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: (01)00884838059450 Serial Numbers: 760012 7742 75043 76137 76005 7830 76202 76134 76038 7871 7825

Product
Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and therapy.
Recalling firm
Philips North America Llc
Quantity
11 units
Distribution
Nationwide Foreign: Austria Belgium Canada China Estonia France Germany Hondorus India Italy Mexico Netherlands Philippines Poland Singapore Spain Switzerland Taiwan
Recall date
November 22, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls