FDAClass IINovember 20, 2013

The Plum A+3 Infusion System is a cassette based recall

The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard…

Why was The Plum A+3 Infusion System is a cassette based recalled?

Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

1) List Number 12348; Serial Numbers: 13740156 , 13740256 , 13740266 , 13740351 , 13740380 , 13740451 , 13740508 , 13740612 , 13740632 , 13740731 , 13740732 , 13740733 , 13740734 , 13740735 , 13740736 , 13740737 , 13740738 , 13740739 , 13740740 , 13740741 , 13740742 , 13740743 , 13740744 , 13740745 , 13740746 , 13740747 , 13740748 , 13740749 , 13740750 , 13740751 , 13740752 , 13740753 , 13740754 , 13740755 , 13740756 , 13740757 , 13740758 , 13740759 , 13740760 , 13740761 , 13740762 , 13740763 , 13740764 , 13740765 , 13740766 , 13740767 , 13740768 , 13740769 , 13740770 , 13740771 , 13740772 , 13740773 , 13740774 , 13740775 , 13740776 , 13740777 , 13740778 , 13740779 , 13740780 , 13740781 , 13740782 , 13740783 , 13740784 , 13740785 , 13740786 , 13740787 , 13740788 , 13740789 , 13740790 , 13740791 , 13740792 , 13740793 , 13740794 , 13740795 , 13740796 , 13740797 , 13740798 , 13740799 , 13740800 , 13740801 , 13740802 , 13740803 , 13740804 , 13740805 , 13740806 , 13740807 , 13740808 , 13740809 , 13740810 , 13740811 , 13740812 , 13740813 , 13740814 , 13740815 , 13740816 , 13740818 , 13740819 , 13740820 , 13740821 , 13740822 , 13740823 , 13740824 , 13740825 , 13740826 , 13740827 , 13740828 , 13740829 , 13740830 , 13740831 , 13740832 , 13740833 , 13740834 , 13740835 , 13740836 , 13740837 , 13740838 , 13740839 , 13740840 , 13740841 , 13740842 , 13740843 , 13740844 , 13740845 , 13740846 , 13740847 , 13740848 , 13740849 , 13740850 , 13740851 , 13740852 , 13740853 , 13740854 , 13740855 , 13740856 , 13740857 , 13740858 , 13740859 , 13740860 , 13740861 , 13740862 , 13740863 , 13740864 , 13740865 , 13740866 , 13740867 , 13740868 , 13740869 , 13740870 , 13740871 , 13740872 , 13740873 , 13740874 , 13740875 , 13740876 , 13740877 , 13740878 , 13740879 , 13740880 , 13740881 , 13740882 , 13740883 , 13740884 , 13740885 , 13740886 , 13740887 , 13740888 , 13740889 , 13740890 , 13740891 , 13740892 , 13740893 , 13740894 , 13740895 , 13740896 , 13740897 , 13740898 , 13740899 , 13740900 , 13740901 , 13740902 , 13740903 , 13740904 , 13740905 , 13740906 , 13740907 , 13740908 , 13740909 , 13740910 , 13740911 , 13740912 , 13740913 , 13740914 , 13740915 , 13740916 , 13740917 , 13740918 , 13740919 , 13740920 , 13740921 , 13740922 , 13740923 , 13740924 , 13740925 , 13740926 , 13740927 , 13740928 , 13740929 , 13740930 , 13740977 , 13740980 , 13740982 , 13740986 , 13740994 , 13741025

Product
The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard…
Recalling firm
Hospira Inc.
Quantity
214 units
Distribution
Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Costa Rica and Dominican Republic.
Recall date
November 20, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)