FDAClass IINovember 6, 2024

AVOXimeter 1000E recall

AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

Why was AVOXimeter 1000E recalled?

Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 10711234530023. Serial Number: 6824

Product
AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.
Recalling firm
Accriva Diagnostics, Inc.
Quantity
1
Distribution
US: AZ
Recall date
November 6, 2024
Classification
Class II
Status
Completed
Agency
FDA (device)
Open FDA enforcement record Latest recalls