Freelite Human Lambda Free Kit recall
Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S
Why was Freelite Human Lambda Free Kit recalled?
Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number: LK018.S UDI Code: 05051700006477 Lot Number: 472776 Model Number:LK018.10S UDI Code: 05051700012201 Lot Number: 472664
- Product
- Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S
- Recalling firm
- The Binding Site Group, Ltd.
- Quantity
- 2,254 kits
- Distribution
- U.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: Canada
- Recall date
- December 1, 2021
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)