FDAClass IINovember 6, 2024

Access Thyroglobulin Antibody II recall

Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access Thyroglobulin Antibody…

Why was Access Thyroglobulin Antibody II recalled?

IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All IFU Part Numbers B14994, B63978, and A38168 affect all product/lots shipped from Jul-2021 to current; UDI-DI 15099590202910.

Product
Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access Thyroglobulin Antibody…
Recalling firm
Beckman Coulter, Inc.
Quantity
465,481 kits
Distribution
Distribution was made nationwide to the U.S., including Puerto Rico. There was government distribution but no military distribution. Foreign distribution was made to Mexico.
Recall date
November 6, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)