Samsung GM60A-40S Intended for use in generating recall
Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.
Why was Samsung GM60A-40S Intended for use in generating recalled?
When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F
- Product
- Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.
- Recalling firm
- NeuroLogica Corporation
- Quantity
- 7 devices
- Distribution
- Distributed to the states of NC, OH, MA, and MO.
- Recall date
- November 18, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)