FDAClass IINovember 18, 2015

GE Healthcare recall

GE Healthcare, SIGNA PET/MR 3.0T. Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission…

Why was GE Healthcare recalled?

When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

00000000UA0064 415723SHMRW 00000000UA0217 415CBMRP N/A 608262PETMR1 MAYOPETMR01 UABPETMR SID TBD SID TBD 262574PETMR1 310974MR1 MRUA0459 309452MR1 082427HWBAY1 M40345211 N/A SID TBD SID TBD SID TBD 00001309FMM076 C001MR01 00000309668MR2 AC001MR02 00000307417MR6 120810MR03

Product
GE Healthcare, SIGNA PET/MR 3.0T. Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission…
Recalling firm
GE Healthcare
Quantity
17
Distribution
Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.
Recall date
November 18, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls