FDAClass IIINovember 6, 2024

17-OH Progesterone ELISA 96 well kit- An enzyme recall

17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in…

Why was 17-OH Progesterone ELISA 96 well kit- An enzyme recalled?

The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: 04048474012927 (DRG Instruments GmbH) lots: 22K024 and 22K034

Product
17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in…
Recalling firm
DRG International, Inc.
Quantity
504 units
Distribution
AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam
Recall date
November 6, 2024
Classification
Class III
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls