FDAClass IINovember 24, 2021

Torque Limiting Adaptor recall

Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation…

Why was Torque Limiting Adaptor recalled?

There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.

Product
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation…
Recalling firm
Laboratoires Bodycad, Inc.
Quantity
10 in the U.S.
Distribution
Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Recall date
November 24, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)