Computed Tomography Systems Revolution CT recall
Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Why was Computed Tomography Systems Revolution CT recalled?
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
- Product
- Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 105 in total
- Distribution
- Worldwide and US Nationwide Distribution
- Recall date
- December 1, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)