BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is recall
BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic…
Why was BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is recalled?
Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI: (10) 1355046 (17) 230131 (30) 0025 (01) 30382904486076 Lot Number: 1355046 Exp. Date: 31JAN2023
- Product
- BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic…
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 13,900 EA
- Distribution
- US Nationwide distribution.
- Recall date
- November 23, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)