FDAClass IIOctober 31, 2018

Camera Arm Drape recall: Sterility Concern

Camera Arm Drape (Box of 20) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system / Distributed by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or nonsurgical instruments/equipment [e.g. robotic arms, microscopes, tables, x ray systems, light handles, etc]. They function as a physical barrier to prevent cross contamination between the instrument / equipment and medical staff and/or to allow the instrument/equipment to enter a hygienic area [e.g. sterile surgical field] in various clinical settings. They are typically made of flexible plastic and shaped to fit the type of instrument / equipment. They are a single use device [not to be reprocessed]

Why was Camera Arm Drape recalled?

The pouches of certain lots of product may have wrinkles along the pouch seal that could result in a channel within the seal. Should there be a channel in the pouch seal it may result in a breach in the sterility of the product.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item 420279-03 Lots # D180437A and D180567

Product
Camera Arm Drape (Box of 20) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system / Distributed by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or nonsurgical instruments/equipment [e.g. robotic arms, microscopes, tables, x ray systems, light handles, etc]. They function as a physical barrier to prevent cross contamination between the instrument / equipment and medical staff and/or to allow the instrument/equipment to enter a hygienic area [e.g. sterile surgical field] in various clinical settings. They are typically made of flexible plastic and shaped to fit the type of instrument / equipment. They are a single use device [not to be reprocessed]
Recalling firm
Ecolab Inc
Quantity
88 cases (1,760 eaches)
Distribution
Worldwide - US Nationwide Distribution in the states of AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, AND WY In the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Iceland, Ireland, Italy, Martinique, Netherlands, Norway, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, and United Kingdom
Recall date
October 31, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)