FDAClass IINovember 6, 2024

ClariTEE Probe- Miniature transesophageal echo recall

ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010…

Why was ClariTEE Probe- Miniature transesophageal echo recalled?

Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI - Primary - 00861589000108 Lot numbers: 20250509 and 20250823

Product
ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010…
Recalling firm
ImaCor Inc.
Quantity
73 units
Distribution
FL ,GA, NJ, TN
Recall date
November 6, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)