FDAClass IIOctober 29, 2025

Fuse ULTRA Foot Plating System Instructions for Use recall

Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L…

Why was Fuse ULTRA Foot Plating System Instructions for Use recalled?

Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All versions of IFU DC-0495 prior to Rev D. All Lots Up to Date 2025/09/10. UDI-DI: B565OS200020L, B565OS200025L, B565OS200120L, B565OS200125, B565OS200125L, B565OS200135, B565OS200324, B565OS200325, B565OS200326, B565OS200400, B565OS200401, B565OS200516, B565OS200518, B565OS200520, B565OS200522, B565OS200616, B565OS200618, B565OS200620, B565OS200622, B565OS216150, B565OS218150, B565OS292110, B565OS292120, B565OS292160, B565OS292200, B565OS326920T, B565OS326922T, B565OS326924T, B565OS326925, B565OS326925T, B565OS326926T, B565OS326928, B565OS326928T, B565OS326930, B565OS326930T, B565OS326932, B565OS326932T, B565OS326934, B565OS326934T, B565OS326936, B565OS326936T, B565OS326938, B565OS326938T, B565OS326940, B565OS326940T, B565OS326945, B565OS326945T, B565OS326950, B565OS326950T, B565OS326955, B565OS326955T, B565OS326960, B565OS326960T, B565OS327920, B565OS327922,B565OS327924, B565OS327926, B565OS327928, B565OS327930, B565OS327932, B565OS327934, B565OS327936, B565OS327938, B565OS327940, B565OS327945, B565OS327950, B565OS327955, B565OS327960, B565OS328017, B565OS328026, B565OS328028C, B565OS421000L, B565OS421000R, B565OS421002L, B565OS421002R, B565OS421004L, B565OS421004R, B565OS421006L, B565OS421006R, B565OS421112, B565OS421116, B565OS421120, B565OS421124, B565OS421130, B565OS421208, B565OS421210, B565OS421212, B565OS421300, B565OS421302, B565OS421304, B565OS421306, B565OS421308, B565OS421406, B565OS421408, B565OS421414, B565OS42150L, B565OS42150M, B565OS42150S, B565OS42150XS, B565OS421512, B565OS421514, B565OS421516, B565OS421710, B565OS421712, B565OS421714, B565OS421716, B565OS421718, B565OS421720, B565OS421722, B565OS421724, B565OS421726, B565OS421728, B565OS421730, B565OS421735, B565OS421740, B565OS421745, B565OS421750, B565OS421814, B565OS421816, B565OS421818, B565OS421820, B565OS421822, B565OS421824, B565OS421826, B565OS421828, B565OS421830, B565OS421835, B565OS421840, B565OS421845, B565OS421850, B565OS422400, B565OS422403, B565OS422404, B565OS422405, B565OS422406, B565OS422407, B565OS422508, B565OS422510, B565OS422512, B565OS422514, B565OS422516, B565OS422518, B565OS422520, B565OS422522, B565OS422524, B565OS422526, B565OS422528, B565OS422530, B565OS422608, B565OS42260LL, B565OS42260LR, B565OS42260ML, B565OS42260MR, B565OS42260SL, B565OS42260SR, B565OS42260XSL, B565OS42260XSR, B565OS422610, B565OS422612, B565OS422614, B565OS422616, B565OS422618, B565OS422620, B565OS422622, B565OS422624, B565OS422626, B565OS422628, B565OS422630, B565OS422702, B565OS422703, B565OS422712, B565OS422714, B565OS422716, B565OS42280, B565OS422803, B565OS422804, B565OS422805, B565OS422807, B565OS422825, B565R327924, B565R327928, B565R327932, B565R327936, B565R327940, B565R327945, B565R327950, B565R421000L, B565R421000R, B565R421710, B565R421712, B565R421714, B565R421716, B565R421718, B565R421720, B565R421722, B565R421724, B565R421726, B565R421728, B565R421730, B565R421735, B565R421740, B565R421814, B565R421816, B565R421818, B565R421820, B565R421822, B565R421824, B565R421826, B565R421828, B565R421830, B565R421835, B565R421840, B565R421845, B565R421850, B565R42260M, B565R42260MR, B565R42260SL, B565R42260SR, B565R42260XSL, B565R42260XSR

Product
Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L…
Recalling firm
CPM Medical Consultants, LLC.
Quantity
3317
Distribution
US Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.
Recall date
October 29, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)