Superion¿ Indirect Decompression System recall
Superion¿ Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm
Why was Superion¿ Indirect Decompression System recalled?
Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model (UPN): 101-9808 UDI-DI (GTIN) Code: 00884662000529 Lot/Batch Numbers: 26176977 Model (UPN): 101-9812 UDI-DI (GTIN) Code: 00884662000543 Lot/Batch Numbers: 26573432 26375758 26815560 26807662 26654632 26928168 26929605 26489700 Model (UPN): 101-9814 UDI-DI (GTIN) Code: 00884662000550 Lot/Batch Numbers: 26366842 26848047 26619510 26323984 26406210 26799951
- Product
- Superion¿ Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Quantity
- 175 devices
- Distribution
- U.S Nationwide Distribution including states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MI, MN, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TX, UT, WI, WV, and WY
- Recall date
- November 23, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)