Bact/alert Virtuo System recall
BACT/ALERT VIRTUO system, B Unit, with Version R2.0 Firmware Product Usage: The BacT/ALERT VIRTU Microbial Detection System is an automated microbial…
Why was Bact/alert Virtuo System recalled?
The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Reference Number 411661, UDI 03573026369774, all serial numbers
- Product
- BACT/ALERT VIRTUO system, B Unit, with Version R2.0 Firmware Product Usage: The BacT/ALERT VIRTU Microbial Detection System is an automated microbial…
- Recalling firm
- bioMerieux, Inc.
- Quantity
- 151 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KS, LA, MI, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SD, TN, TX, UT, VT, WV. Distributed internationally to Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Germany, Spain, France, United Kingdom, Hong Kong, India, Italy, Japan, South Korea, Netherlands, Portugal, Sweden, Singapore, Thailand, South Africa, Saudi Arabia, Philippines, Myanmar, Serbia, Lebanon, United Arab Emirates, Slovenia, Qatar, Costa Rica, Croatia, Panama, Estonia, Israel, Philippines, Malaysia, Brunei Darussalam, Vietnam, Kuwait.
- Recall date
- October 31, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)